Leak Detector Maintenance & Repairs
Our leak detector services range from basic repairs to complete rebuilding of the unit. When we receive the unit we first make an assessment of the problem and then give an estimate for the cost and time to repair. After this is confirmed our technicians proceed with the utmost in care and attention to detail. After the repairs are complete, we retune and calibrate the unit to meet or exceed the original equipment manufacturer's specifications. We also warrant our work for 90 days from shipment. We also offer preventative maintenance programs to fit your needs.
Level 1 (Every 2 Months) WET UNITS
Drain and Flush Mechanical Pump Oil
Fill Pump and Test Performance
Replace Pump Exhaust filter
Check Turbo Pump Start-Up and Performance
Clean Up Vacuum Port Inlet and Outside of Unit
Check all wiring and reseat electrical connectors
Retune Mass Spectrometer with a certified leak standard.
Check Overall Operation of Leak Detector.
Price 855.00
Level 2 (Every 6 Months) WET UNITS
Drain and Flush Mechanical Pump Oil
Fill Pump and Test Performance
Replace Pump Exhaust filter
Disassemble and Clean All Vacuum Components
Clean Up Vacuum Port Manifold and Valves
Clean and Rebuild Mass Spectrometer
Replace Filament and Electron Collector
Check Turbo Pump Start-Up and Performance
Check all wiring and reseat electrical connectors
Retune Mass Spectrometer with a certified leak standard.
Clean Inside Cabinet and Outside of Unit
Check Overall Operation of Leak Detector.
Price 2325.00
Level 3 (Every 12 Months) WET UNITS
Drain and Flush Mechanical Pump Oil
Fill Pump and Test Performance
Replace Pump Exhaust filter
Disassemble and Clean All Vacuum Components
Clean Up Vacuum Port Manifold and Valves
Clean and Rebuild Mass Spectrometer
Replace Filament and Electron Collector
Re-Lubricate Turbo Pump Bearings
Check Turbo Pump Start-Up and Performance
Check all wiring and reseat electrical connectors
Retune Mass Spectrometer with a certified leak standard.
Clean Inside Cabinet and Outside of Unit
Check Overall Operation of Leak Detector.
Price 2975.00
Preventative Maintenance
Our preventative maintenance programs apply to Leybold, Alcatel, Varian, VEECO, and most other brands of leak detectors. Our procedures meet or exceeded all OEM recommendations to maintain your leak detector at optimum performance and maximum sensitivity.
Level 4 (Every 24 Months) WET UNITS
Disassemble and Clean All Vacuum Components
Clean Up Vacuum Port Manifold and Valves
Clean and Rebuild Mass Spectrometer
Replace Filament and Electron Collector
Rebuild Mechanical Pump
Test Mech Pump base pressure and pumping speed.
Rebuild Turbo Pump and Replace Bearings
Check Turbo Pump Start-Up and Performance
Check all wiring and reseat electrical connectors
Retune Mass Spectrometer with a certified leak standard.
Clean Inside Cabinet and Outside of Unit
Check Overall Operation of Leak Detector.
Price 5475.00